HORIZONTAL vs. VERTICAL DILATOMETERS(Technical Note #106)One of the most baffling aspects of selecting a dilatometer is making the choice between a horizontal and vertical unit. Frequently it is cited that the choice is purely a personal preference. In reality, there are substantial differences between the two configurations, each being better suited for a particular task than the other. This note attempts to highlight some of these differences and to give a rule of thumb for rational selection. Anter produces both types of dilatometers, as well as tilted models, and to start with, neither system is better than the other, except with respect to certain specific applications. Horizontal ConfigurationThese instruments have a cantilevered horizontal tube protruding into a horizontal furnace cavity. The pushrod is also cantilevered parallel with the axis of the tube. The sample (cylindrical or rectangular slab) rests on the bottom inside face of the tube, on a cradle or holder, which then rests on the tube. Most frequently the top half of the tube is cut away to make the changing of the sample easy. The end of the tube is closed off either rigidly (such as when the base is actually fused to the tube) or with a plate trapped in grooves. The sample is then slid up against it and held in place by tacking force of the pushrod from the other end. It is clear that the pushrod must exert a reasonable force on the sample to overcome the friction between it and the tube. For a trapped end plate, this pressure must also be sufficient to also keep the plate pressed against the far side of the groove and do not allow it to move at all. Normally this requires a fair amount of force, which in most cases for rigid solids matters little. In one specific application, however, this aspect becomes a major weakness of this configuration. Samples that soften, shrink, or sinter will shorten rather than expand when heated. If the pushrod force is low to prevent indentation into the end of the sample, then it may not be enough to overcome the friction and the sample will not slide to keep the other end pressed against the end plate. As it shrinks a gap may develop between it and the plate. This is totally unpermissible, as the dilatometer just lost track of the process. The end result can even look believable and smooth, but totally wrong. In some dilatometers that use flat leaf spring suspensions for the pushrod, this problem is even further aggravated by the minimal force exerted by the spring. These "frictionless" suspensions refer to the absence of friction in the suspension of the pushrod and not the friction of the sample. For these reasons, it is not advisable to choose a horizontal system for shrinkage or sintering work.
Vertical ConfigurationIn this form, the furnace is either a tube furnace or a pot furnace (bottom of chamber is blanked) and the dilatometer is inserted into it vertically.
Tilted ConfigurationAn infrequently used configuration is the slightly tilted one. Usually around 75° from the horizontal, these essentially vertical devices provide an important advantage. Sometimes it is not easy to stand a sample up on its end or as it is heated, its end may become dished out. This condition allows the sample to "walk" (a movement that is best described by the precession of a spinning top) which then causes large scatter in the data. The same sample in a tilted dilatometer is allowed to lean against the concave inner wall of the tube, keeping it constrained in both directions (x or y) perpendicular to its axis (z). With this, end preparation for samples is considerably relaxed. The friction problem noted for horizontal units is not present, as its own weight is enough to keep the sample in contact with the bottom plate.
home | sitemap | wanted |
what's new | contact us
Unitherm, QuickLine, and FlashLine are Registered Trademarks of Anter Corporation.
|